Journal: Biochemistry and Biophysics Reports
Article Title: Evaluation of a research use only (RUO) TIMP-1 assay for use with the automated Alinity i platform
doi: 10.1016/j.bbrep.2026.102509
Figure Lengend Snippet: Linearity performance of the Alinity i TIMP-1 immunoassay across the analytical measuring interval Observed concentrations for gravimetrically prepared dilution levels are plotted against expected values. All levels demonstrated a deviation from linearity within ±10%, confirming linear response throughout the validated range.
Article Snippet: Here, we describe reformulation of the RUO TIMP-1 immunoassay for migration to the high-throughput automated Alinity i platform (Abbott Laboratories) and assessment of its analytical performance.
Techniques: